Wiring
Stamolys CA71HY
18
Hauser
4.2
Signal connections
4.2.1
Shielding of the analog outputs
The interference suppressor dampens interference from electromagnetic effects on control, power
and signal lines.
After connecting the data transfer cables, clip the interference suppressor (in scope of delivery) onto
the cable cores (not onto the outer insulation of the cable!). You have to route the cable screen
outside around the interference suppressor and connect it to PE (brass screw with bolt) at the top
right in the connection compartment (
18).
!
Note!
With the two-channel version (data cables at analog output 1 and analog output 2), route the cable
cores of the two data cables through the interference suppressor.
4.2.2
One-channel version
a0001746
Fig. 18:
Interference protection of the signal cable
1
Cable shield (at PE in connection compartment!)
2
Signal cable
3
Interference suppressor
4
Cable cores of the signal cable
1
2
3
4
Connection
Designation
Function
Signal inputs
Leak
Liquid has collected in the drip pan
No sample
No sample available, measurement is not started,
display flashes
Signal outputs
AV 1
Alarm value 1 exceeded or undershot
AV 2
Alarm value 2 exceeded or undershot
Fault
Error message, can be retrieved using operation menu
Measurement end
Displays "Measurement finished" (5 s)
Analog output
l-1 channel 1
0 or 4 mA = start of measuring range
20 mA = end of measuring range
Summary of Contents for StamoLys CA 71 HY
Page 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Page 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Page 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Page 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Page 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Page 59: ......
Page 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...