Note!
For the reagents, you need an additional base at max. 35 cm (13.8 inch) under the pumps in the
open version. The dimensions of the reagent bottles are as follows: 90 x 90 x 215 mm (3.54 x 3.54
x 8.46 inch). Depending on the version, the number of bottles varies between 2 and 5.
Furthermore, in these versions the outlet pipe has to be installed to the right beside the analyzer.
Please refer to the sheet enclosed with the Operating Instructions.
The outlet pipe must be secured to a wall in such a way that the incline of the outlet pipes from the
photometer is between 5 and 10 % . You must extend the outlet pipes where necessary.
3.3.2
Connecting the sample line
One-channel version
Two-channel version
• One or two collecting vessels (with or without level monitoring) form part of the scope of delivery,
depending on the version.
• In the two-channel version, level monitoring is only possible for one channel.
• One collecting vessel can be mounted on the housing. A second vessel would have to be made
available near the device.
a0001363
Fig. 6:
Collecting vessel at side wall (optional)
1
Ventilation
2
Sample supply from sample conditioning
3
Collecting vessel
4
Electrical connections
5
Analyzer sample inlet
a0001346
Fig. 7:
Dimensions of collecting vessel and base plate
*
* Dimensions variable, as freely adjustable
6
Sampling for analyzer
7
Sample overflow
8
Analyzer outlet
Collecting vessel (at analyzer outer side, with and without level monitoring)
Connection
Hose ID 3.2 mm (0.13")
Customer collecting vessel
Connection
Hose ID 1.6 mm (0.06")
Max. distance from collecting vessel to analyzer
1 m (3.28 ft)
Max. height difference from collecting vessel to analyzer
0.5 m (1.64 ft)
*
*
*
130
(5.12)
17
(0.67)
10 (0.39)
145 (5.71)
79
(3.1
1)
120
(4.72)
140
(5.51)
115 (4.53)
Ø 34/25 (1.34/0.98)
3 x Ø 28/20 (1.1/0.79)
4 x Ø 6.6 (0.26)
27 (1.06)
60 (2.36)
4 x Ø 5.5/10.4 (0.22/0.41)
Ø 56 (2.2)
Ø 50H7 (1.97H0.28)
M20x1.5
Ø 24 (0.94)
70 (2.76)
100
(3.9)
80(3.15)
mm (inch)
48(1.89)
42(1.65)
70(2.76)
60(2.36)
Summary of Contents for StamoLys CA 71 HY
Page 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Page 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Page 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Page 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Page 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Page 59: ......
Page 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...