Operation
Stamolys CA71HY
30
Hauser
5.3.2
Calibration example
You want to activate an immediate calibration (e.g. after you have replaced the reagents).
You have replaced the reagents and filled the hoses again (no air bubbles). The analyzer is in the
measuring mode.
1.
Hold
H
down until AUTO MEASURING appears.
2.
Use
W
to move through the PARAMETER ENTRY menu and press
F
.
3.
Use
F
to go to the "1. calibration" item.
4.
Now use the
W
,
V
and
F
keys to set a time in the past.
5.
Press
F
to accept the value and then press
H
twice to return to the main menu (AUTO
MEASURING).
6.
Press
F
again. This takes you back to the measuring mode.
The calibration is now carried out automatically.
"
Caution!
After the calibration has finished, the analyzer automatically goes to the measuring mode. You now
have to set the time of the 1st calibration back into the future in order to align the calibration and
rinsing times to each other. Rinsing must be performed 3-4 hours before the next calibration.
Proceed as described above to change the setting for the time of the 1st calibration. After changing
to the measuring mode, the analyzer automatically begins measuring, rinsing and calibrating at the
defined times.
Summary of Contents for StamoLys CA 71 HY
Page 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Page 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Page 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Page 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Page 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Page 59: ......
Page 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...