Stamolys CA71HY
Installation
Hauser
13
6.
Secure the hose boxes in the appropriate pump holders (
12):
Sample pump left, reagent pump right. The direction of flow of the sample and reagent must
be counterclockwise.
7.
Connect the sample supply.
!
Note!
The sample can be obtained as follows:
– Directly or via a reversible flow filter or a cross-flow filter by means of a small pump (rating
approx. 300 ml/min), suitable for clear media, e.g. in the outlet of a sewage treatment plant
– From microfiltration; this is practical for media containing flocculants, e.g. in an activated
sludge basin
– Sample conditioning using ultrafiltration for contaminated media, e.g. from the primary
settling tank.
For questions regarding sample conditioning and its automation, please contact Service or the
sales center responsible for your region.
8.
Connect the tubes from canisters containing reagents, standards and cleaning agents to the
following nozzles:
a0006362
Fig. 11:
Valves and valve hoses
V1-2
1
2
3
Valves
Channel switchover
To sample pump
Connecting hose to valve 1, at rear
4
5
6
7
Outlet hose
Hose, valve 2 at front, standard
Hose, valve 2 at rear, cleaner
Hose, valve 1 at front, sample
a0002206
Fig. 12:
Reagent pump, top view
R1
Reagent 1
R2
Reagent 2 (if present)
R3
Reagent 3 (if present)
Canister
Tube designation (mark)
Sample
Reagent 1
Reagent 2
Standard solution
Cleaner
P
S
R
2
3
5
6
7
V1
V2
1
V3
R1
R2
R3
Summary of Contents for StamoLys CA 71 HY
Page 47: ...Stamolys CA71HY Technical data Endress Hauser 47...
Page 49: ...Stamolys CA71HY Appendix Endress Hauser 49 a0001908 en...
Page 50: ...Appendix Stamolys CA71HY 50 Endress Hauser Two channel version a0001909 en...
Page 51: ...Stamolys CA71HY Appendix Endress Hauser 51 a0001910 en...
Page 56: ...Appendix Stamolys CA71HY 56 Endress Hauser 11 4 Maintenance schedule a0001911 en...
Page 59: ......
Page 60: ...www endress com worldwide BA362C 07 en 07 06 Printed in Germany FM SGML 6 0 DT 51517622...