Maintenance and Service
This chapter provides safety information for the maintenance and service of
the TDLAS Gas Analyzer. For complete instructions refer to the Operating
Instruction → 1.
Cleaning and
Decontamination
To clean the TDLAS gas analyzer exterior
The housing should be cleaned only with a damp cloth to avoid electrostatic
discharge.
Fuse Ratings and
Characteristics
Inside the SS500/SS2000 analyzer electronics enclosure is the electronics
assembly. Fuses are located on the electronics control board → 6–1 and →
Table 6–1 Fuse specifications
Never use vinyl acetate, acetone or other organic solvents to
clean the analyzer housing or labels.
If you need to replace a fuse, use only the same type and
rating of fuse as the original. Refer also to specifications
listed below.
Drawing
Reference
Voltage
Description
Rating
F1
120 VAC
Miniature Fuse, 5 x 20 mm,
Time Delay
250VAC/0.8A
240 VAC
Miniature Fuse, 5 x 20 mm,
Time Delay
250VAC/0.5A
F2
12 VDC
Miniature Fuse, 5 x 20 mm,
Time Delay
250VAC/
3.15A
24 VDC
Miniature Fuse, 5 x 20 mm,
Time Delay
250VAC/1.6A