XA02744C
TDLAS Gas Analyzer
28
Hauser
Ventilation Requirements
Flammable gases or vapors shall not be vented to an area in which the exist
-
ing area classification will be impacted. Flammable gases or vapors shall be
vented to a safe location or to the flare header. For information related to
mitigating risks associated with electrocution, etc → 7.
Equipment Operation
This chapter provides an overview of safety operational instructions for the
SS500XP and SS2000XP TDLAS Gas Analyzers.
Operating Controls
Refer to the appropriate manual for the purchased analyzer to review the
operational controls for the equipment → 1.
Firmware Version
Each SS500XP and SS2000XP TDLAS gas analyzer operates based on its own
firmware version which is listed in the system calibration report and displays
when the analyzer starts up. Detailed operation instructions are provided in
the Description of Device Parameters available for download by model type
on the Hauser website (
www.endress.com
).
Preparing the analyzer for shipment or storage
If the equipment is to be shipped or stored for any length of time, it should
be packed in the original packaging when shipped when shipped from the
factory. If analyzer has been installed and or operated (even for purposes of
a demonstration), the system should first be decontaminated (purged with
an inert gas) before powering down the analyzer.
Storage
The packaged analyzer should be stored in a sheltered environment that is
temperature controlled between -20
C (-4
F) and 50
C (122
F), and should
not be exposed to direct sun, rain, snow, condensing humidity or corrosive
environments.
Process samples may contain hazardous material in
potentially flammable and/or toxic concentrations. Personnel
should have a thorough knowledge and understanding of the
physical properties and safety precautions for the sample
contents before installing the SCS.
Process samples may contain hazardous material in
potentially flammable and/or toxic concentrations.
Personnel should have a thorough knowledge and
understanding of the physical properties of the sample and
prescribed safety precautions before installing, operating or
maintaining the analyzer.