TDLAS Gas Analyzer
XA02744C
Hauser
21
4
SS500XP and SS2000XP TDLAS Gas
Analyzer
The information in this chapter includes technical specifications as well as
safety warnings and considerations for equipment installation and equip
-
ment operation for the SS500XP and SS2000XP TDLAS Gas Analyzers. For
complete instructions refer to the Operating Instruction → 1.
Analyzer Technical
Specifications
All TDLAS Gas Analyzer models come with a set of technical specifications
that outline recommended equipment settings and ratings. This information
is included in the Operating Instruction.
Peripheral devices
For systems equipped with peripheral devices, e.g., probe assemblies, only
devices that meet with Hauser' specifications may be used. Refer to
documentation provided by the manufacturer for instruction on installation,
operation, etc.
Equipment rating
The following equipment rating information is provided by model type for
each TDLAS Gas Analyzer in the
Operating Instruction.
Table 4–1 SS500XP analyzer specifications
Measurement data
Target components
H
O in natural gas
Principle of
measurement
Tunable Diode Laser Absorption Spectroscopy
(TDLAS)
Measurement ranges
0.25 to 20, 0.25 to 50, 0.25 to 100 lbs/MMscf
5 to 422, 5 to 1055, 5 to 2110 ppmv
Repeatability
±1 ppmv or ±1% of reading (whichever is
greater)
Accuracy
±10 ppmv or ±2% of reading (whichever is
greater)
Application Data
Ambient
temperature range
-20 °C to 50 °C (-4 °F to 122 °F)