Soliphant FTM 31 DR
The Soliphant FTM 31 DR with extension tube e.g.
– if mounting is only possible from above
– if the switch point is to be altered using the sliding sleeve (accessory)
– with heavy build-up on the silo wall (mounting from above or from the side)
Take into account the angle of the mound or outflow funnel:
– for calculating the required length of the sensor when ordering a
Soliphant FTM 31 DR
– for calculating the length of the mounting nozzle and mounting point, if you have an
instrument with a specific length.
The Soliphant FTM 31 DR may not be shortened!
Mounting from above:
Mount in the centre of the discharge hopper order to keep the lateral load caused by
discharging material to a minimum; or close to the vessel wall with an extra fastening
near to the fork.
Mounting from the side:
Only order the length of extension tube required as very strong forces can affect the
Soliphant due to outflowing material or when homogenising the material.
Fasten a very long Soliphant extension tube by using a tight-fitting nozzle or fastener.
Maximum length of a mounting nozzle:
Length of the Soliphant from process connection to the tip of the fork minus 7 in
(170 mm).
Check that there is enough space outside the silo for mounting.
a)
b)
c)
d)
BA17
5
Y04
Fig. 4
a) Plate to protect against
inflowing material
b) Sliding sleeve for infinite
adjustment of the switch point
c) sufficient space for mounting
d) Protective hood against
condensation in the housing
FTM 30 DR, FTM 31 DR, FTM 32 DR