Flange connection of the
Soliphant
Advance work:
Important for mounting a narrow nozzle:
– Weld the flange exactly at right angles to the pipe.
– The internal diameter of the pipe should be a min. 1.73 in (44 mm) throughout
its length.
Select a flange gasket suitable for the operating pressure and temperature.
Positioning the fork:
This is only necessary with a laterally mounted Soliphant.
Note!
The function can be affected if material remains lying on the broad side of the fork.
When using the instrument for minimum detection, outflowing material can damage
the fork if the broad side of the fork is uppermost.
It is necessary therefore:
To turn the flange so that the marks are at the top.
This indicates that the narrow side of the fork is upwards and that the material can
flow unhindered and does not remain on the tines.
(min. 44 mm)
= 90°
ø min.1.73 in
= 90°
= 90°
= 90°
FTM
32 S:
max.
5 in
max. 15°
(130
mm)
B
A
175
Y10
Fig. 10
Mounting nozzle and counter
flange welded exactly together
Figure on the extreme right
for FTM 32 DR only
Note!
II
Marks on the flange
BA17
5
Y11
Fig. 11
Position the fork with the
marks at the top!
FTM 30 DR, FTM 31 DR, FTM 32 DR
3 Installation
Hauser
15