Smart System for Aquacultures SSP200B
Electrical connection
Hauser
19
6.4.1
Cable mounting rail
1
3
2
4
A0025171
13
Cable mounting rail and associated function
1
Cable mounting rail
2
Threaded bolt as protective ground connection, central grounding point
3
Additional threaded bolts for ground connections
4
Cable clamps for fixing and grounding the sensor cables
6.4.2
Connecting the Liquiline CM444 supply voltage
L+ N- PE
A
43
42
41
86
85
86
85
98
97
88
87
32
31
32
31
B
A0015872
14
Power supply connection on BASE-E
A
Internal power supply cable
B
Extension power unit
85
85
86
86
41
43
42
L/+ N/–
*
Power
87
88
97
98
31
31
32
32
SD
Display
1
2
Sensor
0/4 ... 20 mA
Sensor supply
+
+
–
+
+
–
1
2
Service
PK
GY
PK
GY
1
2
–
L/+ N/– PE
BN
WH
GN
YE
BN
WH
GN
YE
87
88
97
98
+
–
A
B
+
–
A
B
–
Alarm
HART
Power
(internal)
B
A0015873
15
Overall wiring diagram for BASE-E and extension
power unit (B)
Connecting the supply voltage – power unit 100 to 230 V AC
1. Route the power supply cable through the appropriate cable entry and into the
housing.
2. Connect the protective ground of the power unit to the threaded bolt specially
provided on the cable mounting rail.
3. We recommend you connect the protective ground or ground cable provided onsite to
the mounting bolt. For this purpose, pull the protective ground or the ground cable
through the cable entry and connect it to the threaded bolt on the cable mounting
rail.
4. Connect cable cores L and N to the plug-in terminals of the power unit as shown in
the wiring diagram.
Requirements for protective ground / ground cable
• Onsite fuse 10 A: wire cross-section min. 0.75 mm
2
(18 AWG)
• Onsite fuse 16 A: wire cross-section min. 1.5 mm
2
(14 AWG)