Technical data
RID16
38
Hauser
Humidity
• Condensation permitted as per IEC 60 068-2-33
• Max. rel. humidity: 95% as per IEC 60068-2-30
Degree of protection
IP 67. NEMA 4X.
Shock and vibration resistance
10 to 2000 Hz for 5g as per IEC 60 068-2-6
Electromagnetic compatibility
(EMC)
CE EMC conformity
The device complies with all the requirements of IEC 61326-1:2006 and NAMUR NE21:2007.
This recommendation is a consistent determination as to whether the devices used in laboratories
and in distributed control systems are immune to interference, thus increasing their functional
safety.
Measuring category
Measuring category II as per IEC 61010-1. The measuring category is provided for measuring on
power circuits that are directly connected electrically with the low-voltage network.
Degree of contamination
Pollution degree 2 as per IEC 61010-1.
ESD (electrostatic discharge)
IEC 61000-4-2
6 kV cont., 8 kV air
Electromagnetic fields
IEC 61000-4-3
0.08 to 4 GHz
10 V/m
Burst (fast transients)
IEC 61000-4-4
1 kV
Surge
IEC 61000-4-5
1 kV asym.
Conducted RF
IEC 61000-4-6
0.01 to 80 MHz
10 V