Installation
Prothermo NMT81
30
Hauser
5.11.2 Stilling well method
Prepare the stilling well that is larger than the diameter of the measuring probe when
installing it.
When using an anchor weight, use a pipe that is 100A (4") (JIS, ASME) or larger. If an
anchor weight is not being used in the stilling well method, install the WB probe so that its
end is below the bottom of the stilling well. This will allow the pipe to be filled with liquid.
Ø42 (1.65)
1
≤
150
(5.90)
≤
150
(5.90)
2
1
≥
300 (11.81)
3
A0042754
24
Stilling well. Unit of measurement mm (in)
1
Stilling well
2
Hole (Ø 25 mm (0.98 in))
3
Base plate/datum plate
Stilling well installation procedure
1. Pass the temperature probe and the WB probe through a gasket and insert them
from the installation nozzle at the top of the tank.
2. Use bolts to secure the flange of NMT81 to the installation nozzle on top of the tank.
This completes the procedure for installing a stilling well.
Summary of Contents for Prothermo NMT81
Page 111: ......