Prothermo NMT81
Operability
Hauser
55
7.7.2
NMT81 configuration via NMS8x/NMR8x/NRF81
The following configuration is NMT81-related parameters. For details on the operation of
NMS8x, NMR8x and NRF81, see their respective operating instructions.
The following parameters can be checked from the display accessed through the
Main Menu → Expert → Input/Output → HART device → HART device(s) [MenuName].
Parameter configuration
NMT device config
Configure device?
Total no. element
Bottom point
NMT8NoElementPhase
Water bottom level offset
Update water level
Element setup
Select element
Zero adjust
Element temperature
Element position
Configure device?
Item
Details
Navigation
Expert → Input/Output → HART Device → HART device(s) [MenuName] →
HART device configuration → Config. device? (14728)
Description
Configures the NMT device.
Selection
Yes (Setting can be configured.)
No (No changes in settings. It will be set back to "No" after completion of
configuration.)
Factory setting
No
Additional information
Read access: Operator
Write access: Maintenance
Total no. elements
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Total No. elements (14730)
Description
Displays the total number of elements that can be configured.
Additional information
Read access: Operator
Write access: -