Operability
Prothermo NMT81
56
Hauser
Bottom point
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) [MenuName] →
HART Device configuration → Bottom point (14729)
Description
Displays the bottom-point temperature element.
Input unit
Numerical value (mm)
Factory setting
-
Additional information
Read access: Operator
Write access: Maintenance
Select element
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → Select element (14734)
Description
The element to be configured is manually selected.
Input unit
1-16
Factory setting
1
Additional information
Read access: Operator
Write access: Maintenance
NMT8NoElementInPhase
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → NMT8NoElementInPhase (14756)
Description
Alarm setting for detecting the situation when the none of elements are positioned in
either of vapor, product, nor water phase.
Factory setting
Logbook entry only
Additional information
Read access: Operator
Write access: Maintenance
Water bottom level offset
Item
Details
Navigation
Expert → Input/Output → HART device → HART device(s) → NMT device
configuration → Element configuration → Water bottom level offset (14757)
Description
Enter an offset to adjust the output value of the water bottom probe.
Factory setting
0
Additional information
Read access: Operator
Write access: Maintenance