TEMPERATURE
NMT SET DATA
DIAGNOSTIC CODE
51
Display of current diagnostic code.
0
Display 0 ... 255
480
TOTAL NO. ELEMENT
51
The total number of elements that are mounted in the flexible
tube. This number is determined in accordance with the
specifications provided when the order of the device is
placed.
16
Set
a ... A HEX
482
PREAMBLE NUMBER
51
Display of preambles for HART
®
protocol.
5
Display 2 ... 14 HEX
483
KIND OF INTERVAL
51
Temperature element intervals.
- Equal intervals: 0
- Unequal intervals: 1
If 1 is chosen, then set the element position on matrix from
GVH=460 to GVH=469.
0
Set
0 or 1
485
BOTTOM POINT
51
Height of bottom point.
Only available when equal intervals are selected.
500mm
Set
0.0 ... 99999.9 mm
486
ELEMENT INTERVAL
51
Interval between temperature elements.
Only available when equal intervals are selected.
2000mm
Set
0.0 99999.9 mm
487
TEM. AT SHORT ELE.
51
Temperature indication when element is shorted. This value is
sent to the Proservo only when the error output is "ON" at
GVH=492. When the error output is "OFF", the average
temperature is sent to the Proservo.
-49.5 °C
Set
488
TEM. AT OPEN ELE.
51
Temperature indication when element is opened. This value is
sent to the Proservo only when the error output is "ON" at
GVH=492. When the error output is "OFF", the average
temperature is sent to the Proservo.
359.0 °C
Set
489
Note!