Prothermo NMT 538
Hauser Systems & Gauging
29
QOG 6: Unequal element interval setting more than 11 elements installed
UNEQUAL
SELECT POINT
ELEMENT POSITION
0
655mm
SET DATA NMT
NMT ADJUSTMENT
KIND OF INTERVAL
>GROUP SELECT<: E
>GROUP SELECT<: E
Enter "E" key once, go to
"SELECT POINT" then, select set
number from 0 to 15 by "+" or "-"
key, and "E" to confirm
Set desired element position by
"+" or "-" key, then "E" to confirm.
Enter "E" several times until next
"SELECT POINT" appears, then set
next element position.
Repeat same procedure until the
last element position is set.
"E" key once to confirm
"UNEQUAL INTERVAL" then,
enter "E" several times, return to
"SET DATA NMT"
Enter with "+" or "-" key, going to
"NMT ADJUSTMENT" matrix
Enter "E" 4 times, go to "ELEMENT
POSITION"
Note!
Up to 10 element positions can be set
at 10 matrix of "ELEMENT POSITION"
row."
Caution!!
#1 element is set at "SELECT POINT 0"
; therefore, #16 element will be set at
"SELECT POINT 15".
Choice of "UNEQUAL" allows specific
element position setting
LCD display
Procedure
Remark
5 Operation with the Proservo NMS53x