Prothermo NMT 538
26
Hauser Systems & Gauging
5.1.5
Configuring the NMT538 on the matrix of Proservo NMS53x
To display NMT538 data and configure detailed parameters, the communication between these
2 instruments must be initially established.
QOG 2: Configuring NMT538 on the matrix
This procedure must be completed; other wise, the Proservo will not recognize the connection of
NMT538. The G4 "TEMPERATURE" matrix will be shown only this NMS53x to NMT538
communication is established on the matrix configuration.
SERVICE
OFF
SYSTEM DATA
CALIBRATION
MORE FUNCTION
MORE FUNCTION
MORE FUNCTION
AVERAGE
MATRIX OF
CONNECTION NMT
MATRIX OF
>GROUP SELECT<: E
>GROUP SELECT<: E
>GROUP SELECT<: E
>GROUP SELECT<: E
CONNECTION NMT
Enter "E" key several times
until "MORE FUNCTION" appears
again
Enter "+" key 3 times to display
"SYSTEM DATA", "E" key to confirm
Enter "E" key 3 times to display
"CONNECTION NMT"
Use "+" or "-" key to select temp.
device "AVERAGE"
"E" key once to enter the selection
of Dynamic matrix (G1 to G9)
Enter with "+" key twice, going to
"SERVICE" matrix
Note!
Start at G0 Static Matrix, V4 "MORE
FUNCTION"
Normally, G1: CALIBRATION matrix
to start
Now you are in the
G3 "SERVICE" matrix
Note!
OFF: no temp. device selected
SPOT: connection to 3 wire spot
AVERAGE: connection to NMT
At the "CONNECTION NMT", AVERAGE
must be chosen in order to display
NMT configuration matrix group
G4 "TEMPERATURE"
Move on to the next parameter
setting at G4 "TEMPERATURE"
matrix
Enter "E" to confirm, then enter "E"
several times until "MORE
FUNCTION" appears
LCD display
Procedure
Remark