Status output
Works setting: Error messages
Switch No. 2: OFF
This switched output can optionally be configured for:
•
1) Signalling the flow direction
•
Error messages:
Fault (System error: coil current error, amplifier error DAT error,
EEPROM error, ROM error, RAM error)
Alarm (Process error: measuring range exceeded, v
≥
12.5 m/s)
Supply failure)
The status output acts as a normally closed contact, i.e. in normal operation fault, the
output is closed (transistor conducting, see Fig. 36).
1) The “error messages” are always activated in the fiscal meter.
Note!
The behaviour of the outputs in the event of a fault is described in Chapter 5.1.
➤
System units
Works setting: SI units
Switch No. 3: OFF
SI units: volume in [l, m
3
]
US units: volume in [US gal]
Note!
With the fiscal Promag 31 the SI units are always activated!
Configuration of status
output
Status
Behaviour of open
collector output
Signalling of system and
process errors
System OK
closed
Fault signal/alarm
open
Supply failure
open
«closed»:
open collector
⇒
conducting
«open»:
open collector
⇒
not conducting
b
a028
y
3
3
Fig. 36
4. Operation and Commissioning
Promag 31
30
Hauser