3.5 Cable specifications
Cable specification for remote-mounted version (FS)
Coil cable:
2 x 0.5 mm
2
PVC cable with common screen
Conductor resistance:
≤
37
Ω
/km
Capacitance: core/core, screen earthed
≤
120 pF/m
Signal cable:
3 x 0.38 mm
2
PVC cable with common screen and
separately screened cores.
Conductor resistance:
≤
50
Ω
/km
Capacitance: core/screen
≤
420 pF/m
Cable specification for remote-mounted version (FL)
Coil cable:
2 x 0.5 mm
2
PVC cable with common screen
Conductor resistance:
≤
37
Ω
/km
Capacitance: core/core, screen earthed
≤
120 pF/m
Signal cable:
5 x 0.5 mm
2
PVC cable with common screen
Conductor resistance:
≤
37
Ω
/km
Capacitance: core/core, screen earthed
≤
120 pF/m
Cable specifications for use in areas with severe electrical interference
The Promag 31 measuring equipment fulfils all general requirements for
electromagnetic compatibility (EMC) according to EN 50081 Part 1 and 2 /
EN 50082 Part 1 and 2 when installed in accordance with the NAMUR recommenda-
tions.
Note!
With the remote-mounted version (FS and FL) the signal and coil cables between
sensor and transmitter must always be screened and earthed at both ends.
This is done at the earth terminals inside the connection housing of sensor and
transmitter (see page 26).
Promag 31
3. Electrical Connection
Hauser
27
Summary of Contents for promag 31
Page 4: ...Promag 31 4 Endress Hauser ...
Page 22: ...2 Mounting and Installation Promag 31 22 Endress Hauser ...
Page 28: ...Promag 31 28 Endress Hauser ...
Page 44: ...5 Fault Location and Remedies Promag 31 44 Endress Hauser ...
Page 56: ...6 Technical Data Promag 31 56 Endress Hauser ...
Page 59: ......