Proline t-mass B 150
Installation
Hauser
15
Recommended inlet and outlet runs
20 × DN
5 × DN
5 × DN
25 × DN
20 × DN
5 × DN
5 × DN
5 × DN
40 × DN
5 × DN
50 × DN
20 × DN
1
4
2
6
3
5
A0016943
1
reduction
2
expansion
3
90° elbow or T-piece
4
2 × 90° elbow
5
Control valve
6
2 × 90° elbow (3-dimensional)
Outlet run for pressure or temperature transmitter
PT
2…5 × DN
TT
A0015603
PT Pressure measuring device
TT Temperature measuring device
6.1.2
Requirements from environment and process
Ambient temperature range
Measuring device
–40 to +60
°
C (–40 to +140
°
F)
Local display
–20 to +60
°
C (–4 to +140
°
F), the readability of the display may be impaired at
temperatures outside the temperature range.
►
If operating outdoors: