Proline t-mass 65 PROFIBUS DP
Commissioning
53
5.
Configuration of the Analog Input function blocks 1 to 3:
The measuring device has three Analog Input function block (AI modules), through which the
various measured variables can be cyclically transmitted to the PROFIBUS Master (class 1).
The assignment of a measured variable to the Analog Input function block is shown below
using the example of Analog Input function block 1 (AI module, slot 1).
Using the CHANNEL function, you can determine the measured variable (e.g. mass flow) to
be cyclically transmitted to the PROFIBUS Master (class 1):
a. Select COMMUNICATION
→
BLOCK SELECTION
b. Select the option AI1 - OUT VALUE
c. Select the function CHANNEL
d. Select the option MASS FLOW
6.
Configuration of totalizers 1 to 2:
The measuring device has two totalizers. The following example describes the configuration of
the totalizer using totalizer 1 as an example.
Using the CHANNEL function, you can determine the measured variable (e.g. totalized mass
flow) to be cyclically transmitted to the PROFIBUS Master (class 1) as a totalizer value:
a. Select TOTALIZER
→
SELECT TOTALIZER
b. Select the option TOTALIZER 1
c. Go to the function CHANNEL
d. Select the option MASS FLOW
• Enter the desired unit for the totalizer: TOTALIZER
→
UNIT TOTALIZER
• Configure totalizer status (e.g. totalize): TOTALIZER
→
SET TOTALIZER
→
Option: TOTALIZE
• Set the totalizer mode: TOTALIZER
→
MODE TOTALIZER: Select one of the following options:
– BALANCE (factory setting): calculates the positive and negative flow portions.
– POSITIVE: calculates the positive flow portions.
– NEGATIVE: calculates the positive flow portions.
– HOLD VALUE: The totalizer remains at the last value.
!
Note!
The device can’t calculate negative flow portions.
Possible setting
Measured variable
ID for CHANNEL function
MASS FLOW (factory setting AI function block 1)
277
CORR. VOLUME FLOW (factory setting AI function block 2)
398
TEMPERATURE (factory setting AI function block 3)
285
!
Note!
If, when the PROFIBUS network configuration, the module “AI” was integrated in slot 1, 2 or 3, the measured
variable selected in the CHANNEL function is transmitted cyclically to the PROFIBUS Master (class 1) for the
respective Analog Input function block 1 to 3 (
Possible setting
Totalizer value / measured variable
ID for CHANNEL function
MASS FLOW (factory setting totalizer 1 to 2)
277
CORR. VOLUME FLOW
398
OFF
0
!
Note!
If, when the PROFIBUS network configuration, the module or the function “TOTAL” was integrated in slot 4 or 5,
the measured variable selected in the CHANNEL function is transmitted cyclically to the PROFIBUS Master
(class 1) for the respective totalizer 1 to 2 (