Trouble-shooting
Proline t-mass 65 PROFIBUS DP
80
381
S: FLUIDTEMP.MIN.
!: # 381
0x40
0x41
0x42
UNCERTAIN
O.K.
O.K.
Low
High
Fluid temperature
min.
Cause:
The minimum fluid temperature limit for the
transducer has been exceeded.
Remedy:
Increase the process gas temperature.
"
Achtung!
Caution! In case of severe temperature exposure,
the transducer may be damaged.
382
S: FLUIDTEMP.MAX.
!: # 382
0x40
0x41
0x42
UNCERTAIN
O.K.
O.K.
Low
High
Fluid temperature
max.
Cause:
The minimum fluid temperature limit for the trans-
ducer has been exceeded.
Remedy:
Reduce the process gas temperature.
"
Achtung!
Caution! In case of severe temperature exposure,
the transducer may be damaged.
No. # 5xx
→
Application error
501
S: SW.-UPDATE ACT
!: # 501
0x48
0x49
0x4A
UNCERTAIN
Substitute-Set
O.K.
Low
High
Software upload
active
Cause:
New amplifier or communication (I/O module)
software version is loaded. Currently no other
functions are possible.
Remedy:
Wait until process is finished. The device will
restart automatically.
502
S: UP-/DOWL. ACT.
!: # 502
0x48
0x49
0x4A
UNCERTAIN
Substitute-Set
O.K.
Low
High
Up-/Download
active
Cause:
Up- or downloading the device data via
configuration program. Currently no other
functions are possible.
Remedy:
Wait until process is finished.
561
P: ZERO-ADJ. RUN
!: # 561
0x40
0x41
0x42
UNCERTAIN
O.K.
O.K.
Low
High
Zero adjust running
Cause:
Zero point adjustment function is active.
Remedy:
Wait until process is finished.
No.
Device status message
(local display)
PROFIBUS measured value status
Extended
diagnostic
message in the
PROFIBUS
master
Remedy / spare parts
Qu
a
lit
y
c
o
d
e
(H
EX
)
Me
as
ure
d
va
ri
abl
e
s
tatus
Q
u
al
it
y
s
tatus
Q
u
al
it
y subs
tatus
Li
m
its