General notes
The Quick Setup menu contains the default settings that are adequate for commissioning.
Complex measuring operations on the other hand necessitate additional functions that you can
configure as necessary and customise to suit your process parameters. The function matrix,
therefore, comprises a multiplicity of additional functions which, for the sake of clarity, are arranged
in a number of function groups.
Comply with the following instructions when configuring functions:
• You select functions as described already.
• You can switch off certain functions (OFF). If you do so, related functions in other function groups
will no longer be displayed.
• Certain functions prompt you to confirm your data entries. Press
P
to select “SURE ( YES )” and
press
F
to confirm. This saves your setting or starts a function, as applicable.
• Return to the HOME position is automatic if no key is pressed for 5 minutes.
• Programming mode is disabled automatically if you do not press a key within 60 seconds
following automatic return to the HOME position.
"
Caution!
All functions are described in detail, as is the function matrix itself, in the “Description of Device
Functions” manual, which is a separate part of these Operating Instructions.
!
Note!
• The transmitter continues to measure while data entry is in progress, i.e. the current measured
values are output via the signal outputs in the normal way.
• If the power supply fails all preset and parameterised values remain safely stored in the EEPROM.
• However, some functions may be affected (i. e. data/values not saved) if the power supply is
interupted during editing or operation of those functions. More details are contained in the
“Description of Device Functions” manual, BA114D/06/en/…
5.3.2
Enabling the programming mode
The function matrix can be disabled. Disabling the function matrix rules out the possibility of
inadvertent changes to device functions, numerical values or factory settings. A numerical code
(
factory setting = 65
) has to be entered before settings can be changed.
If you use a code number of your choice, you exclude the possibility of unauthorised persons
accessing data (
→
see the “Description of Device Functions” manual).
Comply with the following instructions when entering codes:
• If programming is disabled and the
P
operating elements are pressed in any function, a prompt
for the code automatically appears on the display.
• If “0” is entered as the customer's code, programming is always enabled.
• Your Hauser representative can be of assistance if you mislay your personal code.
"
Caution!
Changing certain parameters such as all sensor characteristics, for example, influences numerous
functions of the entire measuring system, particularly measuring accuracy.
There is no need to change these parameters under normal circumstances and consequently, they
are protected by a special code known only to the Hauser representative. Please contact
Hauser if you have any questions.
5.3.3
Disabling the programming mode
Programming mode is disabled if you do not press an operating element within 60 seconds following
automatic return to the HOME position.
You can also disable programming in the “ACCESS CODE” function by entering any number (other
than the customer's code).