Proline Promass O 300
Hauser
79
3
1
2
4
4
4
A0033719
46
Options for remote operation via PROFINET network: ring topology
1
Automation system, e.g. Simatic S7 (Siemens)
2
Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication
TCP/IP"
3
Switch, e.g. Scalance X204 (Siemens)
4
Measuring device
Service interface
Via service interface (CDI-RJ45)
A point-to-point connection can be established to configure the device onsite. With the housing
open, the connection is established directly via the service interface (CDI-RJ45) of the device.
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option
NB
: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted in the
cable entry. Therefore the connection to the service interface can be established via an M12
connector without opening the device.
2
3
1
A0027563
47
Connection via service interface (CDI-RJ45)
1
Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated
device Web server or with "FieldCare", "DeviceCare" operating tool with COM DTM "CDI Communication TCP/IP"
or Modbus DTM
2
Standard Ethernet connecting cable with RJ45 connector
3
Service interface (CDI-RJ45) of the measuring device with access to the integrated Web server
Via WLAN interface
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option
G
"4-line, illuminated, graphic display; touch c
WLAN"