Technical data
Proline Promass F 300 EtherNet/IP
250
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the
DAT modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the
previous device data has been exchanged, the new measuring device is ready for
operation again immediately without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are
transferred from the S-DAT in the measuring device and the measuring device is ready
for operation again immediately without any errors
• If exchanging the electronics module (e.g. I/O electronics module): Once the electronics
module has been replaced, the software of the module is compared against the current
device firmware. The module software is upgraded or downgraded where necessary. The
electronics module is available for use immediately afterwards and no compatibility
problems occur.
Manual
Additional parameter data record (complete parameter settings) in the integrated device
memory HistoROM backup for:
• Data backup function
Backup and subsequent restoration of a device configuration in the device memory
HistoROM backup
• Data comparison function
Comparison of the current device configuration with the device configuration saved in
the device memory HistoROM backup
Data transfer
Manual
• Transfer of a device configuration to another device using the export function of the
specific operating tool, e.g. with FieldCare, DeviceCare or Web server: to duplicate the
configuration or to store in an archive (e.g. for backup purposes)
• Transmission of the drivers for system integration via Web server, e.g.:
EDS for EtherNet/IP
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the
Extended HistoROM
application package (order option) is enabled: up to 100
event messages are displayed in the events list along with a time stamp, plain text
description and remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating
tools e.g. DeviceCare, FieldCare or Web server
Data logging
Manual
If the
Extended HistoROM
application package (order option) is enabled:
• Record up to 1
000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g.
FieldCare, DeviceCare or web server