Operation options
Proline Promass F 300 EtherNet/IP
72
Navigation area
If a function is selected in the function bar, the submenus of the function open in the
navigation area. The user can now navigate through the menu structure.
Working area
Depending on the selected function and the related submenus, various actions can be
performed in this area:
• Configuring parameters
• Reading measured values
• Calling up help text
• Starting an upload/download
8.4.6
Disabling the Web server
The Web server of the measuring device can be switched on and off as required using the
Web server functionality
parameter.
Navigation
"Expert" menu → Communication → Web server
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Web server functionality
Switch the Web server on and off.
• Off
• HTML Off
• On
On
Function scope of the "Web server functionality" parameter
Option
Description
Off
• The web server is completely disabled.
• Port 80 is locked.
On
• The complete functionality of the web server is available.
• JavaScript is used.
• The password is transferred in an encrypted state.
• Any change to the password is also transferred in an encrypted state.
Enabling the Web server
If the Web server is disabled it can only be re-enabled with the
Web server functionality
parameter via the following operating options:
• Via local display
• Via Bedientool "FieldCare"
• Via "DeviceCare" operating tool
8.4.7
Logging out
Before logging out, perform a data backup via the
Data management
function
(upload configuration from device) if necessary.
1. Select the
Logout
entry in the function row.
The home page with the Login box appears.