Proline Promass A 100
Installation
Hauser
21
10 mm
13 mm
1
2
3
4
5
6
7 8
A0019746
5
Post retainer mounting kit
1
8 x hexagonal nut M8 × 0.8
2
4 x threaded bolt M8 × 150
3
1 x post retaining plate
4
1 x post securing plate
5
4 x spring washer M8
6
4 x hexagon bolt M6 × 20
7
4 x spring washer M6
8
4 x hexagonal nut M6 × 0.8
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
The zero point obtained in this way is printed on the nameplate of the measuring
device. Calibration takes place under reference conditions . Therefore, a zero point
adjustment in the field is generally not required! )
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with very low flow rates
• Under extreme process or operating conditions, e.g.:
– high process temperature (> 50 °C (122 °F)
– high viscosity (> 100 cSt)
– high process pressure (> 20 bar (290 psi))
6.2
Mounting the measuring device
6.2.1
Required tools
For sensor
For flanges and other process connections: Corresponding mounting tools