Electrical connection
Proline Promass A 100
34
Hauser
Addressing mode ex works
Software addressing; all DIP switches for hardware addressing are set to OFF.
IP address ex works
DHCP server active
For device addressing via software (→ 49)
Setting the address
OFF ON
1
2
3
4
5
6
7
8
9
10
1
2
4
8
16
32
64
128
- Write protection
- Default Ethernet
network settings
IP 192.168.1.212
IP
Address setting
(last octet)
A0017913
1. Set the desired IP address using the corresponding DIP switches on the I/O
electronics module.
Hardware addressing with the configured IP address is enabled after 10 s.
2. Disable software addressing via DIP switch 8 (OFF).
PROFIBUS DP
The address must always be configured for a PROFIBUS DP/PA device. The valid
address range is between 1 and 126. In a PROFIBUS DP/PA network, each address can
only be assigned once. If an address is not configured correctly, the device is not
recognized by the master. All measuring devices are delivered from the factory with the
device address 126 and with the software addressing method.