Proline Promag 50D
Hauser
5
PROFIBUS PA interface
• Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
• Profile version 3.0
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Function blocks: 1 x analog input, 1 x totalizer
• Output data: volume flow, totalizer
• Input data: positive zero return (ON/OFF), totalizer control, value for local display
• Cyclic data transmission compatible with previous model “Promag 33”
• Bus address adjustable via miniature switches or local display (optional) at the measuring device
Signal on alarm
Current output
Failsafe mode can be selected (e.g. in accordance with NAMUR Recommendation NE 43)
Pulse/frequency output
Failsafe mode can be selected
Status output
“Not conductive” in the event of fault or power supply failure
Load
See “Output signal”
Switching output
Status output
• Galvanically isolated
• Max. 30 V DC/250 mA
• Open collector
• Can be configured for: error messages, empty pipe detection (EPD), flow direction, limit values
Low flow cut off
Low flow cut off, switch-on point can be selected as required
Galvanic isolation
All circuits for inputs, outputs and power supply are galvanically isolated from each other.