Proline Promag 50D
8
Hauser
Cable entry
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8 to 12 mm)
• Thread for cable entries, ½" NPT, G ½"
Connecting cable for remote version:
• Cable entry M20 x 1.5 (8 to 12 mm)
• Thread for cable entries, ½" NPT, G ½"
Remote version cable
specifications
Coil cable
• 2 x 0.75 mm
2
(18 AWG) PVC cable with common, braided copper shield (
∅
∼
7 mm)
• Conductor resistance:
≤
37
Ω
/km
• Capacitance core/core, shield grounded:
≤
120 pF/m
• Operating temperature: –20 to +80 °C
• Cable cross-section: max. 2.5 mm
2
(16 AWG)
• Test voltage for cable insulation:
≥
1433 V AC r.m.s. 50/60 Hz or
≥
2026 V DC
Signal cable
• 3 x 0.38 mm
2
(20 AWG) PVC cable with common, braided copper shield (
∅
∼
7 mm) and individual
shielded cores
• Conductor resistance:
≤
50
Ω
/km
• Capacitance core/shield:
≤
420 pF/m
• Operating temperature: –20 to +80 °C
• Cable cross-section: max. 2.5 mm
2
(16 AWG)
a0003194
a
Signal cable
b
Coil current cable
1
Core
2
Core insulation
3
Core shield
4
Core jacket
5
Core reinforcement
6
Cable shield
7
Outer jacket
Operation in zones of severe electrical interference
The measuring device complies with the general safety requirements in accordance with EN 61010-1,
the EMC requirements of IEC/EN 61326 and NAMUR Recommendation NE 21.
"
Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection housing.
Ensure that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible.
Power consumption
Power consumption
• AC: <15 VA (incl. sensor)
• DC: <15 W (incl. sensor)
Switch-on current
• Max 3 A (<5 ms) for 260 V AC
• Max. 13.5 A (<5 ms) for 24 V DC
1
2
3
4
5
6
7
a
b