Proline Promag 50D
18
Hauser
Operating conditions: Process
Medium temperature range
0 to +60 °C
Conductivity
The minimum conductivity is
≥
5
μ
S/cm (
≥
20
μ
S/cm for demineralized water)
!
Note!
In the remote version, the necessary minimum conductivity also depends on the length of the connecting cable
(
→
"Operating conditions: Installation"
→
"Length of connecting cable").
Medium pressure range
(nominal pressure)
• EN 1092-1 (DIN 2501) = PN 16
• ANSI B 16.5 = Class 150
• JIS B2220 = 10 K
Pressure tightness
Measuring tube: 0 mbar abs (0 psi abs) with a fluid temperature of
≤
60 °C
Limiting flow
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor.
The optimum flow velocity is between 2 and 3 m/s. The velocity of flow (v), moreover, has to be matched to
the physical properties of the fluid: v >2 m/s (in the case of fluids that cause buildup, such as wastewater sludge
etc.).
Flow characteristic values
Pressure loss
• No pressure loss if the sensor is installed in a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545
(
→
“Operating conditions: Installation”
→
“Adapters”)
Diameter
Recommended flow
Factory settings
[mm]
[inch]
Min./max. full scale value
(v ~ 0.3 or 10 m/s)
Full scale value current
output
(v ~ 2.5 m/s)
Pulse value
(~ 2 pulses/s)
Low flow cut off
(v ~ 0.04 m/s)
25
1"
9 to 300 dm
3
/min
75 dm
3
/min
0.50 dm
3
1 dm
3
/min
40
1 ½"
25 to 700 dm
3
/min
200 dm
3
/min
1.50 dm
3
3 dm
3
/min
50
2"
35 to 1100 dm
3
/min
300 dm
3
/min
2.50 dm
3
5 dm
3
/min
80
3"
90 to 3000 dm
3
/min
750 dm
3
/min
5.00 dm
3
12 dm
3
/min