Description of device parameters
Proline Promag 300 Modbus RS485
52
Hauser
"Pulse/frequency/switch output 1 to n" submenu
Navigation
Expert → Sensor → Measured val. → Output values → PFS output
1 to n
‣
PFS output 1 to n
Output freq. 1 to n (0471–1 to n)
Pulse output 1 to n (0456–1 to n)
Switch status 1 to n (0461–1 to n)
Output freq. 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Output
freq. 1 to n (0471–1 to n)
Prerequisite
In the
Operating mode
Frequency
option is selected.
Description
Displays the actual value of the output frequency which is currently measured.
User interface
0.0 to 12
500.0 Hz
Pulse output 1 to n
Navigation
Expert → Sensor → Measured val. → Output values → PFS output 1 to n → Pulse
output 1 to n (0456–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected.
Description
Displays the pulse frequency currently output.
User interface
Positive floating-point number
Additional information
Description
• The pulse output is an open collector output.
• This is configured at the factory in such a way that the transistor is conductive for the
duration of the pulse (NO contact) and is safety-oriented.