Description of device parameters
Hauser
141
• Conductivity
• CorrConductivity
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Temperature
• Electronic temp.
Factory setting
Volume flow
Assign diag. beh
Navigation
Expert → Output → Relay output 1 to n → Assign diag. beh (0806–1 to n)
Prerequisite
In the
Relay outp.func.
parameter (→ 139), the
Diag. behavior
option is selected.
Description
Use this function to select the category of the diagnostic events that are displayed for the
relay output.
Selection
• Alarm
• Alarm or warning
• Warning
Factory setting
Alarm
Additional information
Description
If no diagnostic event is pending, the relay output is closed and conductive.
Selection
• Alarm
The relay output signals only diagnostic events in the alarm category.
• Alarm or warning
The relay output signals diagnostic events in the alarm and warning category.
• Warning
The relay output signals only diagnostic events in the warning category.
Assign status
Navigation
Expert → Output → Relay output 1 to n → Assign status (0805–1 to n)
Prerequisite
In the
Relay outp.func.
parameter (→ 139), the
Digital Output
option is selected.
Description
Use this function to select the device status for the relay output.
Selection
• Partial pipe det
• Low flow cut off
*
Visibility depends on order options or device settings