Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
31
Options
Description
Check not done
The device configuration of the HistoROM has not yet been compared to the
backup copy in the device memory.
Dataset incomp.
The backup copy in the device memory is not compatible with the device.
HistoROM
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
3.1.3
"Diagn. handling" submenu
Navigation
Expert → System → Diagn. handling
‣
Diagn. handling
Alarm delay (0651)
‣
Diagn. behavior
Alarm delay
Navigation
Expert → System → Diagn. handling → Alarm delay (0651)
Description
Use this function to enter the time interval until the device generates a diagnostic
message.
The diagnostic message is reset without a time delay.
User entry
0 to 60 s
Factory setting
0 s
Additional information
Result
This setting affects the following diagnostic messages:
• 170 coil resistance
• 832 Electronic temp.
• 833 Electronic temp.
• 834 Process temp.
• 835 Process temp.
• 962 empty pipe