Installation
Oxymax H COS21D
8
Hauser
3
Installation
3.1
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging. Keep the damaged packaging until the
matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents. Keep the damaged products until
the matter has been settled.
• Check that the scope of delivery is complete and agrees with your order and the shipping
documents.
• The packaging material used to store or to transport the product must provide shock
protection and humidity protection. The original packaging offers the best protection. Also,
keep to the approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your sales center responsible.
3.2
Installation conditions
3.2.1
Angle of installation
The sensor must be installed with an angle of inclination of at least 10 ° to the horizontal in an
assembly, support or a suitable process connection. Other angles are not permissible. Do
not
install the sensor overhead.
!
Note!
Make sure you comply with the instructions for installing sensors. You will find them in the
Operating Instructions for the assembly used.
a0005584-en
Fig. 2: Permitted angle of installation
10°
10° Not permissible!
Not permissible!
Permissible angle of installation