Installation
Oxymax H COS21D
10
Hauser
3.3.2
Installing a measuring point
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning
function).
3.
Install and connect the oxygen sensor.
"
Caution!
•
Do not install the sensor suspended from the cable.
• Screw the sensor into the assembly so that the cable is not twisted.
• Mounting or dismounting the sensor: hold the sensor body, only turn the threaded plug-in
head while screwing the sensor in or out of the assembly. Otherwise you could unscrew the
sensor’s membrane cap. It will remain in the assembly resp. in the process then.
• Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
• Select the installation location so that there is easy access for later calibration.
#
Warning!
When using metallic assemblies and installation equipment, comply with national grounding
regulations.