21
3.
Calculate the factor
L
:
T
[°C (°F)]
S
[mg/l=ppm]
T
[°C (°F)]
S
[mg/l=ppm]
T
[°C (°F)]
S
[mg/l=ppm]
T
[°C (°F)]
S
[mg/l=ppm]
0 (32)
14.64
11 (52)
10.99
21 (70)
8.90
31 (88)
7.42
1 (34)
14.23
12 (54)
10.75
22 (72)
8.73
32 (90)
7.30
2 (36)
13.83
13 (55)
10.51
23 (73)
8.57
33 (91)
7.18
3 (37)
13.45
14 (57)
10.28
24 (75)
8.41
34 (93)
7.06
4 (39)
13.09
15 (59)
10.06
25 (77)
8.25
35 (95)
6.94
5 (41)
12.75
16 (61)
9.85
26 (79)
8.11
36 (97)
6.83
6 (43)
12.42
17 (63)
9.64
27 (81)
7.96
37 (99)
6.72
7 (45)
12.11
18 (64)
9.45
28 (82)
7.82
38 (100)
6.61
8 (46)
11.81
19 (66)
9.26
29 (84)
7.69
39 (102)
6.51
9 (48)
11.53
20 (68)
9.08
30 (86)
7.55
40 (104)
6.41
10 (50)
11.25
Altitude
[m (ft)]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
0
1.000
550 (1800)
0.938
1050 (3450)
0.885
1550 (5090)
0.834
50 (160)
0.994
600 (1980)
0.932
1100 (3610)
0.879
1600 (5250)
0.830
100 (330)
0.988
650 (2130)
0.927
1150 (3770)
0.874
1650 (5410)
0.825
150 (490)
0.982
700 (2300)
0.922
1200 (3940)
0.869
1700 (5580)
0.820
200 (660)
0.977
750 (2460)
0.916
1250 (4100)
0.864
1750 (5740)
0.815
250 (820)
0.971
800 (2620)
0.911
1300 (4270)
0.859
1800 (5910)
0.810
300 (980)
0.966
850 (2790)
0.905
1350 (4430)
0.854
1850 (6070)
0.805
350 (1150)
0.960
900 (2950)
0.900
1400 (4600)
0.849
1900 (6230)
0.801
400 (1320)
0.954
950 (3120)
0.895
1450 (4760)
0.844
1950 (6400)
0.796
450 (1480)
0.949
1000 (3300)
0.890
1500 (4920)
0.839
2000 (6560)
0.792
500 (1650)
0.943
L=
relative air pressure during calibration
-------------------------------------------------------------------------
1013 hPa