background image

omnigrad M TR 44, TR 45

8

Hauser

Probe

In TR 45, the measuring probe is constituted by a mineral insulated insert positioned inside the thermowell. 
For its replacement, the insert length (IL) must be chosen depending on the immersion length (L) of the ther-
mowell. Should a spare part be required, please refer to the following table:

In TR 44 the protection well and the sensing part cannot be separated.
The immersion length is available in some standard values or it can be supplied in a "customized" version within
a range (please see the product structures in the last pages of this document). Up to a length of 50 mm the TR
44 may have a probe with 6 mm diameter only.
For special versions of TR 45, where the immersion is shorter than 30 mm, the diameter of the thermowell, as
in the reduced tip, should be 5.3 mm.
The starting material for the wetted parts can be supplied in compliance of Basler Norm 2 (BN2), which
imposes a limited ferrite content and consequently enhances corrosion resistance, on request. In some sensor
configurations, the compliance with the requirements of BN2 can also be assured after the welding and
machining operations, that is in the finished product. The surface roughness (Ra) of the wetted parts is supplied
down to 0.4 mm. Surface roughness below 0.4÷0.5 mm has not been proven to be advantageous in hygienic
applications. Electro-polishing is an electrolytic treatment of the metal surface, which results in it being
cleaned, levelled and passivated.
To improve the heat transfer between the well and the internal sensing insert in the TR 45, a thermo-conduc-
tive compound can be fitted inside the thermowell, on request. The compound should not be requested for
immersion lengths longer than 300 mm, and where it would be used in temperatures of over 200°C.

Fig. 7:

Standard wiring (3 wires)

Fig. 8:

Functional components in TR 45

Sensor

Probe tip

Insert

Insert diam.

Extens. neck

Insert length mm

TR 45

straight

TPR 100

6 mm

82

 

mm

IL = L + 92

TR 45

red./tapered

TPR 100

3 mm

82 mm

IL = L + 92

TR 45

straight

TPR 100

6 mm

E

IL = L + E + 10

TR 45

reduced./tapered

TPR 100

3 mm

E

IL = L + E + 10

1xPt100

3 wires

Red

Red

White

Red

Red

White

121.5

E

L

D

d

R= 3.2

E

L

d

26

D

Sensing
length

Sensing
length

Ø42
Ø33

M4

Ø6

TW

47

h

d

13

01

Summary of Contents for omnigrad M TR 44

Page 1: ...nic process connections supplied as standard other on request Customized immersion length Small probe diameter or reduced tip for fast response time TR 44 Tapered or reduced tip and thermoconductive p...

Page 2: ...de the thermowell and spring loaded underneath it so as to improve the heat transfer As in TR 44 the sensing element Pt100 is positioned close to the tip of the probe The TR 45 thermowell is obtained...

Page 3: ...tions Ambient temperature housing without head mounted transmitter metal housings 40 130 C plastic housings 40 85 C Ambient temperature housing with head mounted transmitter 40 85 C Ambient temperatur...

Page 4: ...length may affect the accuracy of the sensors If it is too low an error might be gen erated in the measurement of the temperature due to the lower temperatures of the process fluid near the walls and...

Page 5: ...revices The use of purely threaded Ingold and metal to metal joints is not recommended by some hygienic design standards i e 3 A Standard 74 02 and Document 8 from EHEDG This is the reason why E H Tem...

Page 6: ...division for hygienic applications because it comes in stain less steel and because of its clean design TA20W BUS type is a round grey coloured head made of aluminium with a clip for the cap closure T...

Page 7: ...es for information regarding availability of welding adaptors The Varivent connections must be used with dedicated Tuchenhagen in line components or tank adaptors For Varivent flanges with small nomin...

Page 8: ...some sensor configurations the compliance with the requirements of BN2 can also be assured after the welding and machining operations that is in the finished product The surface roughness Ra of the we...

Page 9: ...omnigrad MTR 44 TR 45 Endress Hauser 9 Fig 9 TR 44 structure and tip designs E L D d R 3 2 20 3 5 3 Reduced...

Page 10: ...nd calibration the inspection test report consists of a declaration of conformity with the essential points of standard DIN EN 60751 The Factory calibration is carried out in the E H EA European Accre...

Page 11: ...DN 10 15 JD SMS DN 25 1 ND Conical metal to metal connection with G1 2 NE Conical metal to metal connection with M12x1 5 YY Special version Neck dimensions E 30 700 mm and d material stainless steel...

Page 12: ...P TMT181 A PCP 2 wire isolated programmable from to C Q TMT181 B PCP ATEX 2 wire isolated programmable from to C R TMT182 A HART 2 wire isolated programmable from to C T TMT182 B HART ATEX 2 wire iso...

Page 13: ...etal connection with G1 2 YY Special version Neck dimensions E 75 250 mm and d material stainless steel 5 82 mm neck length E 15 mm diameter d 6 mm neck length E to specify 15 mm diameter d 9 mm speci...

Page 14: ...from to C P TMT181 A PCP 2 wire isolated programmable from to C Q TMT181 B PCP ATEX 2 wire isolated programmable from to C R TMT182 A HART 2 wire isolated programmable from to C T TMT182 B HART ATEX 2...

Page 15: ...wire isolated programmable from to C L11 TMT182 A HART 2 wire isolated programmable from to C L21 TMT182 B HART ATEX 2 wire isolated programmable from to C L22 TMT182 C HART FM IS 2 wire isolated prog...

Page 16: ...e paste max temperature 200 C mat nr 60007126 _______________ G1 2 metal to metal weld in adapter for flus mounting of ND process connection material SS 316L 1 4435 provided with leakage detectin port...

Page 17: ...nsmitter iTEMP PCP TMT 181 TI 070R 09 en Temperature head transmitter iTEMP HART TMT 182 TI 078R 09 en Temperature head transmitter iTEMP PA TMT 184 TI 079R 09 en RTD insert for temperature sensors Om...

Page 18: ...uarter Endress Hauser GmbH Co KG Instruments International Colmarer Str 6 79576 Weil am Rhein Germany Tel 49 76 21 9 75 02 Fax 49 76 21 9 75 34 5 www endress com info ii endress com TI 244T 02 en_1204...

Reviews: