Micropilot M FMR231 with FOUNDATION Fieldbus
Trouble-shooting
Hauser
75
9.2.5
Influence of the error codes on the output signal
The following table describes the influence of the error codes on the status of the cyclic output
values as well as on the parameters BLOCK_ERR and XD_ERROR in the Sensor Block.
The output values are linked to the following measuring values:
• Primary Value (PV): level/volume
• Secondary Value (SV): distance between sensor membrane and surface of the material measured
• Third Value (TV): sensor temperature
A164
electronics defect
hardware problem
reset;
if alarm prevails after reset,
exchange electronics
A171
electronics defect
hardware problem
reset;
if alarm prevails after reset,
exchange electronics
A231
sensor 1 defect
check connection
HF module or electronics defective
exchange HF module or electronics
W511
no factory calibration ch1
factory calibration has been deleted
record new factory calibration
A512
recording of mapping please
wait
mapping active
wait some seconds until alarm
disappears
A601
linearisation ch1 curve not
monotone
linearisation not monotonously increasing
correct linearisation table
W611
less than 2 linearisation
points for channel 1
number of entered linearisation points < 2 correct linearisation table
W621
simulation ch. 1 on
simulation mode is active
switch off simulation mode
E641
no usable echo
channel 1
check calibr.
echo lost due to application conditions or
built up on antenna
check installation;
optimize orientation of antenna;
clean antenna (cf. OM)
E651
level in safety distance - risk
of overspill
level in safety distance
alarm will disappear as soon as level
leaves safety distance;
E671
linearisation ch1 not
complete, not usable
linearisation table is in edit mode
activate linearisation table
Code
Description
Possible cause
Remedy
Code
PV Status
SV Status
PV Substatus
SV Substatus
TV Status
TV Substatus
BLOCK_ER
XD_ERROR
A102
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
W103
Uncertain
Non specific
GOOD
Non specific
Other
Unspecified Err
A106
BAD
Device Failure
BAD
Device Failure
Other
Unspecified Err
A110
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance
Electronic Failure
A111
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A113
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A114
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
A115
BAD
Device Failure
BAD
Device Failure
Device needs maintenance now
Unspecified Err
A116
BAD
Device Failure
BAD
Device Failure
Device needs maintenance now
Unspecified Err
A121
BAD
Device Failure
BAD
Device Failure
Memory Failure/
Device needs maintenance now
Electronic Failure
W153
Uncertain
Non specific
GOOD
Non specific
Power up
No Error
Summary of Contents for Micropilot M FMR231
Page 95: ...Micropilot M FMR231 with FOUNDATION Fieldbus Technical data Endress Hauser 95 ...
Page 112: ...Micropilot M FMR231 with FOUNDATION Fieldbus Index 112 Endress Hauser ...
Page 113: ......
Page 114: ...www endress com worldwide BA229F 00 ae 06 07 52009950 CCS FM SGML 6 0 ProMoDo 52008600 ...