Micropilot FMR67B HART
Diagnostics and troubleshooting
Hauser
49
Reset device
parameter
For details see the "Description of device parameters" documentation.
11.10.2 Resetting the device via keys on the electronic insert
Reset the password
Display
on
off
II
15 s
2×
I
I
I
I
I
A0050210
14
Sequence - resetting the password
Deleting/resetting the password
1. Press operating key I three times.
The Reset Password function is started, the LED flashes.
2. Press operating key I once within 15 s.
The password is reset, the LED flashes briefly.
If operating key I is not pressed within 15 s, the action is canceled and the LED is no longer
lit.
Resetting the device to the factory setting
12 s
50 ms
Display
on
o
ff
I
II
I
II
A0050009
15
Operating keys on the electronic insert
Resetting the device to the factory setting
‣
Press operating key I and operating key II simultaneously for at least 12 s.
Device data are reset to the factory setting, the LED flashes briefly.
11.11 Device information
All the device information is contained in the
Information
submenu.
Navigation: System → Information