Micropilot FMR67B HART
Installation
Hauser
17
The maximum nozzle length H
max
depends on the nozzle diameter D
D
H
max
H
max
øD
80 to 100 mm (3.2 to 4 in)
1
700 mm (67 in)
100 to 150 mm (4 to 6 in)
2
100 mm (83 in)
≥ 150 mm (6 in)
3
200 mm (126 in)
In the case of longer nozzles, reduced measuring performance must be expected.
Please note the following:
• The end of the nozzle must be smooth and free from burrs.
• The edge of the nozzle should be rounded.
• Mapping must be performed.
• Please contact the manufacturer's support department for applications with nozzles
that are higher than indicated in the table.
5.3.3
Drip-off antenna PTFE 50 mm (2 in)
Information concerning threaded connections
• When screwing in, turn by the hex bolt only.
• Tool: open-ended wrench 55 mm
• Maximum permissible torque: 50 Nm (36 lbf ft)
Information about the mounting nozzle
The maximum nozzle length
H
max
depends on the nozzle diameter
D
.
Maximum nozzle length H
max
as a function of the nozzle diameter D
D
H
max
H
øD
max
50 to 80 mm (2 to 3.2 in)
750 mm (30 in)
80 to 100 mm (3.2 to 4 in)
1
150 mm (46 in)
100 to 150 mm (4 to 6 in)
1
450 mm (58 in)
≥150 mm (6 in)
2
200 mm (88 in)
In the case of longer nozzles, reduced measuring performance must be expected.
Please note the following:
• The end of the nozzle must be smooth and free from burrs.
• The edge of the nozzle should be rounded.
• Mapping must be performed.
• Please contact the manufacturer's support department for applications with nozzles
that are higher than indicated in the table.