Diagnostics and troubleshooting
Micropilot FMR51, FMR52 FOUNDATION Fieldbus
96
Hauser
Error
Possible cause
Remdy
If the surface is not calm (e.g.
filling, emptying, agitator
running), the measured value
jumps sporadically to a
higher level
Signal is weakened by the rough
surface - the interference echoes are
sometimes stronger.
• Carry out tank mapping (
Confirm
distance
• Select
=
Process vessel with agitator
.
• Increase integration time (Expert
→ Sensor → Distance → Integration
time)
• Optimize orientation of the antenna
• If necessary, select a better mounting
position and/or larger antenna.
During filling/emptying the
measxured value jumps
downwards
Multiple echoes
• Check
Tank type
parameter
• If possible, do not select central
installation position.
• If appropriate, use a stilling well.
Error message F941 or S941
"Echo lost"
Level echo is too weak.
• Check
Medium group
parameter
• If necessary, select a more detailed
setting in
Medium property
• Optimize alignment of antenna
• If necessary, select a better
installation position and/or larger
antenna.
Level echo suppressed.
Delete mapping and record it again.
Device displays a level when
the tank is empty.
Interference echo
Carry out mapping over entire
measuring range when the tank is
empty (
Confirm distance
parameter
Wrong slope of the level in
the entire measuring range
Wrong tank type selected.
Set
Tank type
correctly.