Micropilot FMR51, FMR52 FOUNDATION Fieldbus
Operating menu
Hauser
127
Tube diameter
Navigation
Setup → Tube diameter
Prerequisite
=
Bypass / pipe
Description
Specify diameter of bypass or stilling well.
User entry
0 to 9.999 m
Factory setting
0 m
Medium group
Navigation
Setup → Medium group
Prerequisite
Liquid
Description
Select medium group.
Selection
• Others
• Water based (DC >= 4)
Factory setting
Others
Additional information
This parameter roughly specifies the dielectric constant (DC) of the medium. For a more
detailed definition of the DC use the
Medium property
parameter (→ 138).
The
Medium group
parameter presets the
Medium property
follows:
Medium group
Others
Unknown
Water based (DC >= 4)
DC 4 ... 7
The
Medium property
parameter can be changed at a later point of time. However,
when doing so, the
Medium group
parameter retains its value. Only the
Medium
property
parameter is relevant for the signal evaluation.
The measuring range may be reduced for small dielectric constants. For details refer
to the Technical Information (TI) of the respective device.
Empty calibration
Navigation
Setup → Empty calibr.
Description
Specify the distance E between the process connection and the minimum level (0%). This
defines the starting point of the measuring range.