Technical data
Memosens COS81D
38
Hauser
13
Technical data
13.1 Input
Measured variables
Dissolved oxygen [mg/l, µg/l, ppm, ppb or %SAT or hPa]
Oxygen (gaseous) [hPa or %Vol]
Temperature [˚C, ˚F]
Measuring ranges
Measuring ranges apply for 25 ˚C (77 ˚F) and 1013 hPa (15 psi)
c-shaped
u-shaped
0.004 to 26 mg/l
0.05 to 285 % SAT
0.1 to 600 hPa
0.004 to 30 mg/l
0.05 to 330 % SAT
0.1 to 700 hPa
13.2 Performance characteristics
Response time
From air to nitrogen at reference operating conditions:
• t
90
: < 10 s
• t
98
: < 20 s
Reference operating
conditions
Reference temperature:
25 °C (77 °F)
Reference pressure:
1013 hPa (15 psi)
Maximum measured
error
At 25 °C (77 °F)
Measured value
[mg/l]
Maximum measured error
[mg/l]
Measured value
[hPa]
Maximum measured error
[hPa]
0.04
±0.008
1
±0.2
0.8
±0.017
20
±0.4
9.1
±0.1
210
±2
26
±0.5
600
±12
13.3 Environment
Ambient temperature
range
–5 to +100 ˚C (23 to 212 ˚F)
Storage temperature
-25 to 50 °C (77 to 120 °F)
at 95% relative air humidity, not condensating
Degree of protection
IP68 (10 m (33 ft) water column at 25 ˚C (77 ˚F) over 28 days)
1)
In accordance with IEC 60746-1 at rated operating conditions