Memosens COS81D
Technical data
Hauser
39
IP69K (test in accordance with DIN 40050-9)
13.4 Process
Process temperature
Sensor
General
Oxygen measurement
COS81D-****1* (EPDM)
-10 to +140 °C (15 to 280 °F)
COS81D-****3* (FFKM)
0 to +140 °C (32 to 280 °F)
COS81D-**C*** (c-shaped)
0 to 60 °C (32 to 140 °F)
COS81D-**U*** (u-shaped)
0 to 80 °C (32 to 175 °F)
Process pressure
0.02 to 13 bar (0 to 190 psi) abs.
Temperature-pressure
ratings
p, abs.
T
[bar]
[°C]
6
13
140
110
[psi]
188
87
T
[°F]
230
284
-10
14
0
32
A0032287-EN
Chemical resistance
NOTICE
Halogen-containing solvents, ketones and toluene
Halogen-containing solvents (dichloromethane, chloroform), ketones (e.g. acetone,
pentanone) and toluene have a cross-sensitive effect and result in decreased measured
values or, at worst, in the complete failure of the sensor!
‣
Use the sensor only in media that are free from halogens, ketones and toluene.
CIP compatibility
Yes
SIP compatibility
Yes, max. 140 °C (284 °F)
Autoclavability
Yes, max. 140 °C (284 °F)
13.5 Mechanical construction
Design
The sensor's spot cap can have either a c-shaped or u-shaped design.