Troubleshooting
Levelflex M FMP40 with HART/4...20 mA
68
Hauser
9.2
System error messages
Code
Description
Possible cause
Remedy
A102
Checksum error
general reset & new calibr.
required
Device was switched off before data could
be stored;
EMC problem
E
2
PROM defective
Reset
Avoid EMC problems;
if alarm prevails after reset,
exchange electronics
W103
Initializing - please wait
E
2
PROM storage not yet finished
Wait a few seconds;
If error prevails, exchange
electronics
A106
Downloading - please wait
Processing data download
Wait until warning disappears after
the download procedure
A110
Checksum error
general reset & new calibr.
required
Device was switched off before the data
were saved
EMC problem
E
2
PROM defective
Reset
Avoid EMC problems;
if alarm prevails after reset,
exchange electronics
A111
Electronics defective
RAM defective
Reset
If alarm prevails after reset,
exchange electronics
A113
Electronics defective
ROM defective
Reset
If alarm prevails after reset,
exchange electronics
A114
Electronics defective
E
2
PROM defective
Reset
If alarm prevails after reset,
exchange electronics
A115
Electronics defective
General hardware problem
Reset
If alarm prevails after reset,
exchange electronics
A116
Download error
Repeat download
Checksum of stored data not correct
Restart download of data
A121
Electronics defective
No factory calibration present
E
2
PROM cleared
Contact service
W153
Initializing - please wait
Initialization of electronics
Wait a few seconds; if warning
prevails, switch off device and
switch it on again
A160
Checksum error
general reset & new calibr.
required
Device was switched off before the data
were saved
EMC problem
E
2
PROM defective
Reset
Avoid EMC problems;
If alarm prevails after reset,
exchange electronics
A164
Electronics defective
Hardware problem
Reset
If alarm prevails after reset,
exchange electronics
A171
Electronics defective
Hardware problem
Reset
If alarm prevails after reset,
exchange electronics
A221
Probe pulse deviation from
average values
HF module or cable between HF module
and electronics defective
Check contacts on HF module
If fault cannot be eliminated:
Replace HF module
A261
HF cable defective
HF cable defective or HF connector
removed
Check HF connector, replace cable
if defective
W275
Offset too high
Temperature at the electronics too high or
HF module defective
Check temperature, replace
HF module if defective
W512
Recording of mapping -
please wait
Mapping active
Wait a few seconds until alarm
disappears
W601
Linearization ch1 curve not
monotone
Linearization not monotone increasing
Correct table
Summary of Contents for Levelflex M FMP40
Page 4: ...Brief overview Levelflex M FMP40 with HART 4 20 mA 4 Endress Hauser...
Page 87: ...Levelflex M FMP40 with HART 4 20 mA Appendix Endress Hauser 87...
Page 88: ...Appendix Levelflex M FMP40 with HART 4 20 mA 88 Endress Hauser...
Page 90: ...Levelflex M FMP40 with HART 4 20 mA Index 90 Endress Hauser...
Page 92: ...www endress com worldwide BA363F 00 en 03 09 No 71074941 CCS FM SGML 6 0 ProMoDo 71074941...