Installation
iTEMP TMT71
14
Hauser
Fig. B
Mounting in a field housing
1
Field housing cover
2
Mounting screws with springs
3
Head transmitter
5
Field housing
120 mm
(4.72 in)
140 mm
(5.51 in)
21 mm
(0.83 in)
!6.5 mm
(0.25 in)
A0024604
4
Dimensions of angle bracket for wall mount (complete wall mounting set available as accessory)
Procedure for mounting in a field housing, Fig. B:
1. Open the cover (1) of the field housing (4).
2. Guide the mounting screws (2) through the lateral bores in the head transmitter (3).
3. Screw the head transmitter to the field housing.
4. After wiring, close the field housing cover (1) → 18again.
Fig. C
Mounting on DIN rail (DIN rail as per IEC 60715)
1
Mounting screws with springs
2
Head transmitter
3
Circlips
4
DIN rail clip
5
DIN rail
Procedure for mounting on a DIN rail, Fig. C:
1. Press the DIN rail clip (4) onto the DIN rail (5) until it engages with a click.
2. Fit the mounting springs on the mounting screws (1) and guide the screws through
the side boreholes of the head transmitter (2). Then fix both mounting screws with
the snap rings (3).