Source container FQG61, FQG62
Hauser
39
Connecting the pneumatic actuator
NOTICE
This section only applies to source containers with a pneumatic actuator. (In the product
structure: feature 020, versions K, L, M or N)
‣
L
CAUTION
The pneumatic actuator may only be put into operation once the radiation source container has
been mounted
‣
Compressed air connection
3
1
5
6
7
2
4
9
9
8
A0018411
1
Terminal box for connection of the proximity switches
2
Cable glands for cable diameter 5 to 10 mm (0.2 to 0.4 in)
3
Throttle check valve for compressed air connection
4
Connection for potential equalization
5
Vent filter
6
Connecting cable of proximity switch for "AUS/OFF" switch position
7
Connecting cable of proximity switch for "EIN/ON" switch position
8
Proximity switch for "EIN/ON" switch position
9
Proximity switch for "AUS/OFF" switch position
The compressed air line is connected to the throttle check valve.
L
CAUTION
The throttle check valve is set at the factory and secured with thread locking fluid.
‣
Do not change the setting of the throttle check valve.
Connecting the proximity
switches
1 2 3 4
+ - + -
A0034001
19
Terminal assignment
1
Proximity switch for "EIN/ON" switch position, positive lead (brown)
2
Proximity switch for "EIN/ON" switch position, negative lead (blue)
3
Proximity switch for "AUS/OFF" switch position, positive lead (brown)
4
Proximity switch for "AUS/OFF" switch position, negative lead (blue)