Source container FQG61, FQG62
Hauser
45
FQG61/FQG62; order code
020, option C
Switching ON the radiation
L
CAUTION
Safety instructions for switching ON the radiation
‣
Before switching on the radiation, make sure that no-one is in the radiation zone (or inside the
product vessel).
‣
The radiation may only be switched ON by properly instructed staff.
Switching ON the radiation
L
WARNING
Rotating the insert over the locking pin leads to the removal position of the source insert.
‣
Do not press the sealed locking pin (A)
1. Remove the padlock.
2. Release the locking screw (optional).
AUS
OFF
A
2.
1.
A0018419
3. Turn the swivel bracket 180° in the counterclockwise direction. The current switch state is
indicated by the sign that is visible ("EIN - ON" or "AUS - OFF"). The other sign is covered by the
swivel bracket.
EIN
ON
AUS
OFF
3.
A0018420
4. Secure the "ON" switch position by hooking the padlock into the position provided.