Source container FQG61, FQG62
12
Hauser
dose rate diagram for the real loading and version can be ordered with order code 580 "Test,
Certificate".
For the allocation of the device version, see the Product Configurator on the Hauser
website: www.endress.com → Select your country → Products → Select measuring technology,
software or components → Select the product (picklists: measurement method, product family
etc.) → Device support (right-hand column): Configure the selected product → The Product
Configurator for the selected product opens
Dose rate diagrams for
60
Co
90°
90°
30°
30°
60°
60°
120°
120°
300°
300°
330°
330°
0°
0°
270°
270°
210°
210°
240°
240°
180°
180°
150°
150°
A
B
60% 100%
60% 100%
A0018270
A
FQG61
B
FQG62
Option in order code 100
"Prepared for source
activity"
FQG61
Activity in MBq
FQG62
Activity in MBq
FQG61
Max. value (100%)
in μSv/h
FQG62
Max. value (100%)
in μSv/h
AA
3.7
3.7
0.04
0.01
AB
7.4
7.4
0.08
0.02
AC
18.5
18.5
0.21
0.05
AD
37
37
0.42
0.10
AE
74
74
0.85
0.20
AF
111
111
1.27
0.30
AG
185
185
2.12
0.50
AH
370
370
4.24
1.01
AK
740
740
8.49
2.02
AL
-
1110
-
3.03
AM
-
1850
-
5.04
AN
-
3700
-
10.09